Steam Quality Test
iCALsg is a SQ Test service provider for qualifying plant/utility/clean/pure steam generators, steam distribution systems and steam supplies to autoclaves in accordance with cGMP (Orange Guide), HTM 01, HTM 2010: 1994, HTM 2031: 1997, EN 285: 2006, DIN 58950, ISO 14937:2000, AAMI ST79, ISPE Baseline Guide for Steam and Water and PDA Technical Reports No.1 & No.48 - TR1 & TR48
Type of Service
- Supply of Test Equipments
- On‐site testing and Report generation for each Point of use (POU)
Steam Quality Test & Acceptance Criteria
Non-Condensable Gas (NCG) Test
Acceptance Criteria: ≤ 3.5%
Dryness value test
Acceptance Criteria: ≥ 0.95
Superheat test
Acceptance Criteria: ≤ 25 ⁰C
Temperature difference in Steam Supply Pipe during Dryness and Super Heat test
Acceptance Criteria: ± 3⁰C
Our Equipment Specification
Non-condensable gas test kit | 0-20% N/C per 100ml condensed |
Dryness test | Full range |
Superheat test | Full range |
4 mm steam supply tube | Max 14 BarG Steam |
8mm Coolant supply tube | Max 3 BarG at 20°C |
Condensing unit (Steam side) | Max 5 BarG at 160°C |
Condensing unit (cooling side) | Max 4 BarG at 20°C |
Pitot Tubes | Max 6 BarG at 165°C |